Conclusion:
This study confirmed the necessity and utility of CTC during monopolar
PEF ablation of cardiac tissues via an investigational LAF catheter in a
swine model. Using LI as an indicator, consistent CTC resulted in
effective PEF ablation across a range of cardiac targets suggesting that
contact, per se, is more important than the degree of contact. These
results diverge from previous research using focal catheters, where
increased contact resulted in increased lesion size and indicate LAF
catheter require unique research and workflows. Lesion characteristics
were optimized and reproducible across a wide range of LI changes and,
notably, acute lesion characteristics were preserved chronically. These
results suggest an important role for contact detection in the guided
delivery of PEF energy.