Reproducibility and repeatability of identifying latest electrical activation during mapping of coronary sinus branches in CRT recipients
  • +2
  • Mads Kronborg,
  • Charlotte Stephansen,
  • Jens Kristensen,
  • Christian Gerdes,
  • Jens Nielsen
Mads Kronborg
Aarhus University Hospital Skejby

Corresponding Author:[email protected]

Author Profile
Charlotte Stephansen
Aarhus University Hospital Skejby
Author Profile
Jens Kristensen
Aarhus University Hospital Skejby
Author Profile
Christian Gerdes
Aarhus University Hospital Skejby
Author Profile
Jens Nielsen
Aarhus University Hospital Skejby
Author Profile

Abstract

Introduction: Studies have shown an association between outcome in cardiac resynchronization therapy (CRT) and longer interventricular delay at the site of the left ventricular (LV) lead. Targeted LV lead placement at the latest electrically activated segment increases LV function further as compared with standard treatment. We aimed to determine reproducibility and repeatability of identifying latest electrically activated segment during mapping of all available CS branches in patients receiving CRT. Methods: We included 35 patients who underwent CRT implantation with protocolled mapping guided LV lead implantation aiming for the site of latest electrical activation. Three different doctors experienced in electrophysiology and implantation of CRT devices independently measured time interval from the local bipolar right ventricular (RV) electrogram (EGM) to the local unipolar LV EGM at all mapped sites (RV-LV). The segment with the latest electrical activation was defined as target segment (TS) and the CS tributary containing TS was defined as target vein (TV). Weighted Kappa statistics with 95% confidence intervals were computed to assess intra-and inter-observer agreement for TS and TV. Results: We mapped 258 segments within 131 veins. Weighted kappa values for repeatability were 0.85 (0.81-0.89) for TS and 0.92 (0.89-0.93) for TV, and weighted kappa values of inter-observer agreement ranged from 0.70 (0.61-0.73) to 0.80 (0.76-0.83) for TS and 0.73 (0.64-0.78) to 0.86 (0.83-0.89) for TV among all three observers. Conclusion: The reproducibility and repeatability of identifying latest electrically activated segment during mapping of all available CS branches in patients receiving CRT ranges from good to very good.
16 Jun 2020Submitted to Journal of Cardiovascular Electrophysiology
16 Jun 2020Submission Checks Completed
16 Jun 2020Assigned to Editor
17 Jun 2020Reviewer(s) Assigned
29 Jul 2020Review(s) Completed, Editorial Evaluation Pending
02 Aug 2020Editorial Decision: Revise Minor
11 Aug 20201st Revision Received
11 Aug 2020Submission Checks Completed
11 Aug 2020Assigned to Editor
11 Aug 2020Reviewer(s) Assigned
19 Aug 2020Review(s) Completed, Editorial Evaluation Pending
19 Aug 2020Editorial Decision: Accept
Nov 2020Published in Journal of Cardiovascular Electrophysiology volume 31 issue 11 on pages 2940-2947. 10.1111/jce.14729