Occlusion Following Deployment Of MANTA VCD After TAVR
- Mohammed Hassan,
- Timothy Mixon,
- William Gray,
- Ramachandra Reddy
Abstract
We present a complication following deployment of the MANTA VCD device
following a TAVR procedure which resulted in occlusion of the common
femoral artery. This was addressed by ballooning the site from the
contralateral side which re-established flow. We believe this is the
first report to address this kind of complication and may prove useful
as more of the MANTA devices are being used in multiple procedures.31 Jul 2020Submitted to Journal of Cardiac Surgery 03 Aug 2020Submission Checks Completed
03 Aug 2020Assigned to Editor
04 Aug 2020Reviewer(s) Assigned
08 Aug 2020Review(s) Completed, Editorial Evaluation Pending
10 Aug 2020Editorial Decision: Accept
Nov 2020Published in Journal of Cardiac Surgery volume 35 issue 11 on pages 3148-3149. 10.1111/jocs.14971