Ferrowax microvalve-utilized centrifugal microfluidic system for a fully
automated serial dilution
- Soohyun Kim,
- Jayeon Song,
- RaKyeom Kim,
- Nae Yoon Lee,
- Min Hwan Kim,
- Hyun Gyu Park
Jayeon Song
Korea Advanced Institute of Science and Technology
Author ProfileRaKyeom Kim
Korea Advanced Institute of Science and Technology
Author ProfileHyun Gyu Park
Korea Advanced Institute of Science and Technology (KAIST)
Author ProfileAbstract
We herein describe an ingenious centrifugal microfluidic system to
accomplish a fully automated serial dilution. The liquid flow on the
disc was automatically regulated by utilizing ferrowax microvalves
systematically integrated into the channels within the specially
designed metering structure. By opening the differently positioned
microvalves through irradiation of IR laser to allow metering, the same
amount of diluent was serially eluted to the dilution chamber from the
same diluent chamber. After dilution, the diluted samples were
automatically delivered to the respective final product chambers by
appropriately opening or closing the microvalves in the connecting
channels, followed by rotating the disc. Based on this unique design
principle, six consecutive two-fold and ten-fold dilutions were
successfully achieved, yielding excellent accuracy in a wide dynamic
range up to six orders of magnitude. Very importantly, the overall
serial dilution process, including the diluent addition, mixing, and
product transfer steps, was completed very rapidly within five minutes,
due to the minimized procedures enabled by the automated actuation of
the ferrowax microvalves at the rationally designed positions. The
centrifugal microfluidic system would serve as a powerful elemental tool
to realize the fully automated diagnostic microsystem involving the
serial dilution process.12 Mar 2021Submitted to Biotechnology Journal 12 Mar 2021Submission Checks Completed
12 Mar 2021Assigned to Editor
12 Mar 2021Reviewer(s) Assigned
27 Apr 2021Editorial Decision: Revise Major
05 Jul 20211st Revision Received
07 Jul 2021Submission Checks Completed
07 Jul 2021Assigned to Editor
07 Jul 2021Reviewer(s) Assigned
19 Aug 2021Editorial Decision: Revise Minor
26 Aug 20212nd Revision Received
26 Aug 2021Submission Checks Completed
26 Aug 2021Assigned to Editor
26 Aug 2021Reviewer(s) Assigned
01 Sep 2021Editorial Decision: Revise Minor
06 Sep 20213rd Revision Received
06 Sep 2021Submission Checks Completed
06 Sep 2021Assigned to Editor
06 Sep 2021Reviewer(s) Assigned
07 Sep 2021Editorial Decision: Accept